Lesson 6.3 · 8 min
§ 7 Reprocessing instruments and sterilisation
Open in the coursewith narrated slides, a checklist to tick off and a quiz
At a glance
- What reprocessing is. Cleaning, disinfection and sterilisation for reuse, so that the device poses no risk.
- Separate clean and unclean. A zone for clean and unclean, separated by space or by time. One person is named responsible for reprocessing.
- Steam sterilisation. A steriliser to EN 285 or EN ISO 13060; at least 3 minutes at 134 degrees or 15 minutes at 121 degrees; pressure and temperature documented.
- Pack, store, log. Sterile goods heat-sealed or in containers, stored dry and dust-free, and labelled. Endoscope reprocessing is logged.
- Also when others reprocess. The required records apply to external reprocessing too. When in doubt, single-use; and ask for a reprocessing statement when buying.
In detail
Principles for instruments and materials
- Only CE-marked medical devices are used, for the purpose the manufacturer set Hygiene-V § 20(1).
- Single-use instruments are disposed of after one use; reusable ones are used, cleaned, disinfected and, depending on use, sterilised as the manufacturer specifies § 20(2), (4).
- Clean and unclean material are kept apart, in storage too; used material always counts as unclean. There are instructions for reprocessing that the staff concerned demonstrably know § 21.
The reprocessing procedure
- Reprocessing covers cleaning, disinfection and sterilisation with the related steps, and checking and restoring technical and functional safety. The reprocessed device must pose no risk, from infection or changed function for example § 23(1)–(2).
- For devices used low-germ or sterile there is a zone for clean and unclean. It can be separated by space or by time, within one room too § 23(3).
- A doctor or a member of staff is named responsible for reprocessing § 23(4).
- Cleaning and disinfection are manual or in washer-disinfectors § 23(6). Manual procedures follow documented standard operating procedures with tested agents suited to the device § 23(8).
- Sterilisation uses a steam steriliser to ÖN EN 285 or EN ISO 13060: at least 3 minutes at 134 °C or 15 minutes at 121 °C. The actual pressure and temperature are documented, printed or stored electronically. Instruments with cavities need a pre-vacuum process § 23(7).
Packing and storage
Sterile goods are packed with a heat sealer or in sterile containers, stored dry and dust-free, and labelled § 24.
Wound care and endoscopes
- For instruments used to change dressings and remove stitches, manual cleaning and disinfection with a tested product is enough; for wound care, the wound kit is also sterilised § 26.
- Endoscopes are pre-cleaned right after use following a work instruction. Endoscopes without a working channel are reprocessed by machine with records, or by hand with instructions and a daily signature; endoscopes with a working channel in a washer-disinfector, by hand only in an emergency; endoscopes for sterile areas are sterilised as well § 27. Every endoscope reprocessing is logged, following the template in Annex 1 § 29. Contaminated endoscopes and instruments go in separately marked containers § 30.
Records and doubtful cases
- The legally required record of reprocessing applies even when an external contractor reprocesses for you, wholly or partly § 5(2)(f).
- If in doubt whether something can be reprocessed, use single-use products. When buying a device, ask for a statement on how it is reprocessed § 28.
- External validation of reprocessing is needed only if an authority demands it with reason, unless § 68 of the Medical Devices Act says otherwise § 25.
The template forms of Annex 1 and the further annexes, on risk assessment and reprocessing for example, can be downloaded from the Medical Chamber, ÖQMED and www.arzthygiene.at § 31.
Checklist
- One person is named responsible for reprocessing.
- Clean and unclean are separated, by space or by time.
- Manual procedures have documented standard operating procedures.
- The steriliser meets EN 285 or EN ISO 13060; pressure and temperature of every cycle are documented.
- Sterile goods are packed, labelled, and stored dry and dust-free.
- Endoscope reprocessing is logged; external reprocessing is documented.
Quiz
What values does the Hygiene-V 2014 give for steam sterilisation?
- 10 minutes at 100 °C
- At least 3 minutes at 134 °C or 15 minutes at 121 °C, with documented process parameters
- One hour at 80 °C
- None; only the manufacturer's instructions apply
Show the answer
The answer is B: At least 3 minutes at 134 °C or 15 minutes at 121 °C, with documented process parameters. Hygiene-V § 23(7): a steam steriliser to ÖN EN 285 or EN ISO 13060, at least 3 minutes at 134 °C or 15 minutes at 121 °C, with the actual pressure and temperature documented.
Sources
This lesson's statements rest on:
- Hygiene Regulation of the Austrian Medical Chamber (Hygiene-V 2014), consolidated after the 3rd amendment of 2022
- Health quality portal (BIQG): From self-evaluation to certificate, with the annexes of the Hygiene Regulation and the equipment lists
- Quality Assurance Regulation 2024 (QS-VO 2024), Federal Legal Information System (RIS), version of 3 October 2026
Not legal advice. What counts is the text of the QS-VO 2024 in the Federal Legal Information System (version of 3 October 2026) and the Austrian Medical Chamber's Hygiene Regulation 2014. Not an offer of ÖQMED, the BIQG or the Medical Chamber.